Résumé
We report the case of obstructive sleep apnea in a 19-year-old, otherwise healthy male presenting with persistent daytime sleepiness and nonrestorative sleep after velo- and uvuloplasty. An individually tailored prototype of an orally inserted pharyngeal stenting device was proposed in the framework of a first clinical feasibility trial. The noninvasive, easily self-administered device is mounted on a simple inferior dental guard. Baseline total apnea-hypopnea index (AHI) was 15.5 and 24.4 per hour of rapid eye movement (REM) sleep. With the device, total AHI dropped to 6.7 per hour (56.8% reduction) and 1.4 per hour of REM (94.3% reduction). Recorded sleep efficiency during treatment was excellent at 96.5%. Laryngoscope, 129:1945–1948, 2019.
| langue originale | Anglais |
|---|---|
| Pages (de - à) | 1945-1948 |
| Nombre de pages | 4 |
| journal | Laryngoscope |
| Volume | 129 |
| Numéro de publication | 8 |
| Les DOIs | |
| état | Publié - août 2019 |