TY - GEN
T1 - A 60μW 60 nV/√Hz readout front-end for portable biopotential acquisition systems
AU - Yazicloglu, Refet Firat
AU - Merken, Patrick
AU - Puers, Robert
AU - Van Hoof, Chris
PY - 2006
Y1 - 2006
N2 - A biopotential readout front-end can be configured to extract EEG, ECG, and EMG signals and draws 20μA from 3V. AC coupling of chopped amplifiers results in an input-referred noise of 60nWVHz and CMRR of 120dB at 1kHz. The immunity of the CMRR to electrode offset voltages is improved with an active input stage and 110dB CMRR is achieved at 100Hz with 50mV electrode offset.
AB - A biopotential readout front-end can be configured to extract EEG, ECG, and EMG signals and draws 20μA from 3V. AC coupling of chopped amplifiers results in an input-referred noise of 60nWVHz and CMRR of 120dB at 1kHz. The immunity of the CMRR to electrode offset voltages is improved with an active input stage and 110dB CMRR is achieved at 100Hz with 50mV electrode offset.
UR - http://www.scopus.com/inward/record.url?scp=34247338024&partnerID=8YFLogxK
M3 - Conference contribution
AN - SCOPUS:34247338024
SN - 1424400791
SN - 9781424400799
T3 - Digest of Technical Papers - IEEE International Solid-State Circuits Conference
SP - 56+45
BT - 2006 IEEE International Solid-State Circuits Conference, ISSCC - Digest of Technical Papers
T2 - 2006 IEEE International Solid-State Circuits Conference, ISSCC
Y2 - 6 February 2006 through 9 February 2006
ER -