Abstract
We report the case of obstructive sleep apnea in a 19-year-old, otherwise healthy male presenting with persistent daytime sleepiness and nonrestorative sleep after velo- and uvuloplasty. An individually tailored prototype of an orally inserted pharyngeal stenting device was proposed in the framework of a first clinical feasibility trial. The noninvasive, easily self-administered device is mounted on a simple inferior dental guard. Baseline total apnea-hypopnea index (AHI) was 15.5 and 24.4 per hour of rapid eye movement (REM) sleep. With the device, total AHI dropped to 6.7 per hour (56.8% reduction) and 1.4 per hour of REM (94.3% reduction). Recorded sleep efficiency during treatment was excellent at 96.5%. Laryngoscope, 129:1945–1948, 2019.